arrowELECTROSURGE 400D
Digital 400w Surgical Diathermy(Tur Model)
KEY FEATURES:
  • Pure Cut– Non-modulated smooth Rf-currents with a frequency of 330 khz, produce smooth cuts with less electro-hemostatics effect, means fast and better post-operative healing.
  • Coagulation– RF of 330 kHz modulated by 650 kHz to 25 kHz.
  • Blend– Carrier mixture cut and coag currents variable in steps from 0 to 9.
  • Highest Safety Standards – The state of art electronic circuitry continuously monitors disconnection in the Inactive Electrode and gives audible and visual alarms and switches off output instantaneously.
  • Controls – Simple colour coded controls enable Surgeons to choose between Foot Switch or Hand Switch and Monopolar (or)Bipolar selection. Also desired output mode and power can be selected in just 2 button touch.
TECHNICAL SPECIFICATIONS:
  Modes : 2 Channels
bullet Monopolar Cut       : 400 Watts @ 300 ohms
330KHz sinusoid         
bullet Monopolar Coag      : 330 Watts @  300 ohms
330KHz sinusoid
Burst Width -16 micro Sec
Repetition Rate – 32KHz
bullet Interference Frequency : Base – 0 Hz –100 Hz
Spectrum – 0 Hz – 150 Hz
bullet Bipolar Coagulator       : 90 watts 500KHz sinusoid
bullet Alarms      : Audio visual alarms with output cutoff for Inactive cable, Active cable, Patient plate.
OTHER SPECIFICATIONS:
bullet OPERATING Voltage : 220 V AC, 50 HZ
bullet MAX POWER CONSUMPTION : 600 Watts    
bullet L x B x H : 48 x 40 x 14 cms
bullet Weight : 18 Kgs.
STANDARD ACCESSORIES:
bullet Active cable    
bullet In-active cable      
bullet Needles ( set of 4 needles)     
bullet Surgical Handle ( Autoclavable)     
bullet Silicon Patient Plate    
bullet Foot Switch     
bullet Bipolar forceps with cable     
bullet Disposible Handswitch